Читать книгу Small Animal Laparoscopy and Thoracoscopy онлайн
221 страница из 232
Waveforms
There are three basic types of waveforms generated in electrosurgery: Cutting waveform, coagulation waveform, and blended waveforms (ssss1). Cutting waveforms are continuous waveforms, and coagulation and blended are intermittent waveforms.
The continuous cutting waveform uses less peak voltage at a similar power setting to the intermittent waveform, resulting in less lateral thermal tissue damage. With pure cut, the amplitude of the continuous waveform is the same. This induces a localized effect on the tissues, creating high tissue temperatures (>100 °C), vaporization of the interstitial fluid, and tissue separation with little hemostasis. Heat is absorbed by water released from the cells, which minimizes thermal damage and provides minimal coagulation. Power settings for the cut waveform are often between 50 and 80 W [4].
The intermittent waveform of the coagulation waveform produces a high current density delivered in pulses. This waveform has a higher voltage than the continuous waveforms and delivers the electrical charge deeper into the tissues. The pauses between the pulses lead to decreased tissue heating, resulting in coagulation rather than cutting. Coagulation is achieved with a power setting between 30 and 50 W [4]. Coagulation can either be performed using desiccation or fulguration. Whereas fulguration uses a noncontact technique to control diffuse hemorrhage, desiccation is a contact technique used for local bleeding. Fulguration uses an intermittent waveform, which allows for proteins to melt and recongeal, forming a coagulum. Some ESUs also contain a spray mode, which is useful for oozing capillary beds. Spray mode does not penetrate as deeply into tissues and can be used in more delicate tissues. Desiccation or a contact technique leads to less heat production than fulguration. A coagulum is formed by tissues drying out and proteins melting.